Sorensen XT Series (已停產)

產品名稱:(42-60W) Linear Programmable DC power in a quarter-rack package產品型號:Sorensen XT SeriesSorensen - XT Series (42-60W)Linear Programmable DC power in a quarter-rack packageProduct OverviewThe Sorensen XT provides 60 watts of programmable linear DC power in a quarter-rack package. W

產品介紹

產品名稱:(42-60W) Linear Programmable DC power in a quarter-rack package

產品型號:Sorensen XT Series

Sorensen - XT Series (42-60W)
Linear Programmable DC power in a quarter-rack package

Product Overview 
The Sorensen XT provides 60 watts of programmable linear DC power in a quarter-rack package. With seven models to choose from, XT is the smart choice for both benchtop and system applications where up to 60 watts of programmable, clean DC power is required. XT is also ideal for OEM applications where a wide adjustment of output voltage or current is required for up to 60 watts in a compact package.

The Sorensen XT is a preferred source for system development and test. It is available in singles, duals, triples and quads in a single package for benchtop use. For systems applications, multiple units can also be rack mounted in one to four unit configurations for up to four independent 60–watt outputs. XT can also be combined in a rack with the 300–watt HPD or 500–watt XPD.

Key Features

● Very low noise and ripple
● Excellent line/load regulation
● Fast transient response
● Constant voltage or constant current mode
● LabWindows® drivers
● CE, CSA approvals

Specifications

Output : Voltage and Current
ModelOutput VoltageOutput CurrentOutput Power
XT 7-60-7 V0-6 A42 W
XT 15-40-15 V0-4 A60 W
XT 20-30-20 V0-3 A60 W
XT 30-20-30 V0-2 A60 W
XT 60-10-60 V0-1 A60 W
XT 120-0.50-120 V0-0.5 A60 W
XT 250-0.250-250 V0-0.25 A60 W
ModelLine Regulation² VoltageLine Regulation² CurrentLoad Regulation³ VoltageLoad Regulation³ Current
XT 7-62 mV6.25 mA2 mV6.25 mA
XT 15-42 mV4.25 mA2 mV4.25 mA
XT 20-32 mV3.25 mA2 mV3.25 mA
XT 30-22 mV2.25 mA2 mV2.25 mA
XT 60-12 mV1.25 mA2 mV1.25 mA
XT 120-0.52 mV0.75 mA2 mV0.75 mA
XT 250-0.253 mV0.50 mA3 mV0.50 mA
ModelOutput Noise & Ripple
(20 Hz – 20 MHz) Voltage
Output Noise & Ripple
(20 Hz – 20 MHz) Current
Meter Accuracy Voltage
(1% of Vmax + 1 count)
Meter Accuracy Current
(1% of Vmax + 1 count)
XT 7-6< 1 mVrms< 1 mArms0.08 V0.07 A
XT 15-4< 1 mVrms< 1 mArms0.25 V0.05 A
XT 20-3< 1 mVrms< 1 mArms0.3 V0.04 A
XT 30-2< 1 mVrms< 1 mArms0.4 V0.03 A
XT 60-1< 1 mVrms< 1 mArms0.7 V0.02 A
XT 120-0.5< 1 mVrms< 1 mArms2.2 V0.006 A
XT 250-0.25< 1 mVrms< 1 mArms3.5 V0.003 A
ModelDrift (8 hours)[4]
Voltage (0.02% of Vmax)
Drift (8 hours)[4]
Current (0.03% of Imax)
Temperature Coefficient[5]
Voltage (0.015% of Vmax/°C)
Temperature Coefficient[5]
Current (0.02% of Imax/°C)
XT 7-61.4 mV1.8 mA1.05 mV1.2 mA
XT 15-43 mV1.2 mA2.25 mV0.8 mA
XT 20-34 mV0.9 mA3 mV0.6 mA
XT 30-26 mV0.6 mA4.5 mV0.4 mA
XT 60-112 mV0.3 mA9 mV0.2 mA
XT 120-0.524 mV0.15 mA18 mV0.1 mA
XT 250-0.2550 mV0.075 mA37.5 mV0.05 mA
1. Specifications indicate typical performance at 25° C ± 5°C, nominal line input of 115 Vac.
2. For input voltage variation over the AC input voltage range, with constant rated load.
3. For 0-100% load variation, with constant nominal line voltage.
4. Maximum drift over 8 hours with constant line, load, and temperature, after 30-minute warm-up.
5. Change in output per °C change in ambient temperature, with constant line and load.
6. Apply accuracy specifications according to the following voltage program accuracy example: Set a model 15-4 power supply to 10 V. The expected result will be within the range of 10 V ± 20 mV ± 0.1% of the set voltage of 10 V.
XT 60 W Internal Interface Specifications with RS232 or GPIB Interface Installed [1,6]
ModelProgram Accuracy 
Voltage (mV)
Program Accuracy 
Current (mA)
Program Accuracy 
OVP (mV)
XT 7-6
10 (±0.1%)
110 (±0.15%)
70
XT 15-4
20 (±0.1%)
70 (±0.15%)
150
XT 20-3
20 (±0.1%)
50 (±0.15%)
200
XT 30-2
30 (±0.15%)
40 (±0.15%)
300
XT 60-1
200 (±0.15%)
26 (±0.2%)
200
XT 120-0.5
400 (±0.15%)
13 (±0.2%)
1200
XT 250-0.25
800 (±0.15%)
7 (±0.2%)
2400
ModelReadback Accuracy Voltage (mV)Readback Accuracy Current (mA)
XT 7-6
10 (±0.15%)
110 (±0.15%)
XT 15-4
10 (±0.15%)
70 (±0.15%)
XT 20-3
10 (±0.15%)
50 (±0.15%)
XT 30-2
15 (±0.15%)
40 (±0.15%)
XT 60-1
35 (±0.15%)
26 (±0.2%)
XT 120-0.5
70 (±0.15%)
13 (±0.2%)
XT 250-0.25
140 (±0.15%)
7 (±0.2%)
Input
Operational AC 
Input Voltage
Standard: 115 Vac ±10%. 57-63 Hz; Optional: 110/220/230/240 Vac ±10%, 47-63 Hz
AC Input CurrentSingle Unit: 1.2 A, Dual Unit: 2.4 A, Triple Unit: 3.6 A, Quad Unit: 4.8 A, at 115 Vac
General
Voltage Mode Transient Response Time< 100 μs recovery to 0.05% band, ±50% load change in the range of 25% to 100% 
of the rated load
Front Panel Voltage and Current Control10-turn voltage and 1-turn current potentiometers (10-turn current optional)
Front Panel Voltage Control Resolution0.02% of maximum voltage
AC Input Connector TypeIEC 320 connector
ApprovalsCSA certified to CSA C22.2 No. 107.1. Meets USA EMC standard FCC Part 15B Class A; meets Canadian EMC standard: ICES-001, Class A (Models up to and including 120 V)
Analog Programming (with optional APG interface installed)
Remote On/Off 
and Interlock
2 to 25 Vdc high. < 0.8 Vdc low. User-selectable.
Remote Analog Programming Option0-10 Vdc for 0-100% or rated voltage or current ±0.1%, 0-10 kΩ for 0-100% of 
rated voltage or current ±0.1%
Remote Monitoring0-10 Vdc for 0-100% of rated voltage or current ±0.1%
Over Voltage Protection 
Trip Range
3 V to full output +10%
Tracking Accuracy±1% for series operation
Environmental
Operating Temperature0 to 30°C for full rated output. Above 30°C, derate output linearly to zero at 70°C
Storage Temperature- 55 to 85°C
Humidity RangeUp to 80% RH, non-condensing
Physical
DimensionsWidth: 4.2” (109.2 mm)
Height: 5.2” (134.7 mm)
Depth: 11.7” (297 mm)
WeightApproximately 7.7 lb (3.5 kg)
Specifications subject to change without notice.